Displaying report 1-1 of 1.
Reports until 15:21, Tuesday 02 September 2025
H1 SQZ
camilla.compton@LIGO.ORG - posted 15:21, Tuesday 02 September 2025 - last comment - 10:07, Thursday 04 September 2025(86692)
SQZ NLG Messurements at Different ISS AOM Setpoints

Tried to measure NLG (following 76542) at different ISS setpoints (by changing the SHG waveplate to adjust how much power is incident on the AOM), with the same OPO TRANS power. Because in 86363 we were confused that the NLG increased significantly when we realigned Pump fiber, with the same OPO Trans setpoint.

This was confusing and I'd like to repeat next week. I initially thought we were getting pump depletion so decreased the seed power, but later noticed the ISS was just unable to keep up and the control voltage was dropping to -5 for the low values of ISS setpoint, see attached, it could hold in LOCKED_CLF_DUAL with 80uW OPO_TRANS fine but not with the SEED beam.

OPO Setpoint ISS Setpoint when locked on Amplified Unamplified NLG Notes
CLF DUAL SEED Max Min UnAmp Dark
80 4.9 3.2 0.192924 0.002389 0.0079612 -2.1e-5 24.2  
80 2.9 2.9 0.045037     -2.2e-5 39 ?(using below unamp) Pump deplation?  Reduced SEED power from 0.7 to 0.3 to keep it locked on SEED (still didn't work first time).
80 6.3 6.3 0.0430464 0.000546 0.0011063 -2.2e-5 38 ? Unamp signal decreased from SEED power change.
80 4.8 4.7 0.0454944 0.0005378 0.0018533 -2.7e-5 24.2  
80 2.9 -5 ?           Noticed ISS Controlmon at -5. 
80 3.5 3.5 0.0452317 0.00054049 0.0018636 -2.4e-5 24.0  
80 4.95   0.04523       24.0 Leaving here.
 
Images attached to this report
Comments related to this report
camilla.compton@LIGO.ORG - 10:07, Thursday 04 September 2025 (86729)

 Repeated while Corey was relocking today. We had one strange measurement when the ISS setpoint at 3.3V where the un-amplified signal was much lower but when later repeating at 3.1V, we didn't see this.

OPO Setpoint ISS Setpoint when locked on Amplified Unamplified NLG
CLF DUAL SEED Max Min UnAmp Dark
80 5.0 5.0 0.042961 0.00051437 0.00176935 -2.1e-5 24.0
80 2.8 ISS Setpoint dropped to -5,  so OPO TRANS not at 80uW
80 3.3 3.2 0.043273 0.0005247 0.00106713 -2.1e-5 39.7
80 6.2 6.2 0.041031 0.00051677 0.0017658 -2.1e-5 23.0
80 6.5 6.4 0.0409285 0.00051425 0.00175578 -2.1e-5 23.0
80 3.6 3.5 0.0429806 0.00051216 0.00175932 -2.7e-5 24.1
80 3.1 3.0 0.0431685 0.000516166 0.0017296 -2.7e-5 24.5
80 2.8 ISS Setpoint dropped to -5,  so OPO TRANS not at 80uW
80 4.9 4.8 0.0429243 0.0005134 0.0017627 -2.7e-7 24.0
Displaying report 1-1 of 1.