(In the attached 1 week trend, [-7 day, -2 day] range is before the power outage and [-1day, 0 day] is after the long high power lock post power outage. Ignore [-2 day, -1 day] because the ISS 2nd loop was messing with the AOM during that time.)
ISS QPD (top right) shows a clearly measurable shift in the alignment mostly in DX (YAW). The QPD is downstream of the PMC, the calibration is supposed to be in mm so nominally it's just 5.4 um but there's a focusing lens in front of it and unfortunately I don't know the focal length nor the distance between the lens and the QPD for now, so the only thing I can say is that the alignment changed downstream of PMC in YAW.
BEA (Bull's Eye Sensor, top left) also shows an alignment shift upstream of PMC. Input matrix of this one is funny, don't read too much into the magnitude of the change. I'm not even sure if it was mostly in PIT or in YAW. (The beam size, 2nd from the top on the left column, also shows a change but I won't trust that because that output is supposed to change when there's an alignment change.)