Reports until 10:38, Thursday 25 September 2025
H1 AOS
camilla.compton@LIGO.ORG - posted 10:38, Thursday 25 September 2025 (87139)
SRM Offsets still look good at 5kHz HOM
We checked what the 5kHZ HOM peak looked like in +4dB Mid-SQZ, after adjusting the matrix for OMC-DCPD_524K_A2 as in 85937.  Checking as before our PSL power issues,  we adjusted SRM offsets in 86414 to minimize it. The 5kHz HOM peak still looked minimized as there was no 5kHz peak, see black trace in attached dtt. All other traces apart from "Today MidSQZ" black and "Today FDS" cyan are old. 
Dtt saved /ligo/home/camilla.compton/Documents/sqz/templates/dtt/20250925_5kHz.xml 
Images attached to this report