Displaying report 1-1 of 1.
Reports until 16:23, Friday 26 September 2025
H1 SUS (SUS)
ibrahim.abouelfettouh@LIGO.ORG - posted 16:23, Friday 26 September 2025 (87164)
Weekly In-Lock SUS Charge Measurement - FAMIS 28424

Closes FAMIS 28424, last checked in alog 86887.

Measurements did not seem to run last week (and couldn't find an alog for it), but did this week.

ITMX did not run this week - just like the previous check on 09/09.

I didn't capture a screenshot in time but I recall the same message from the last check: 

"Cannot calculate beta/beta2 because some measurements failed or have insufficient coherence!
Cannot calculate alpha/gamma because some measurements failed or have insufficient coherence!
Something went wrong with analysis, skipping ITMX_13_Hz_1440859844"

 

Plots attached.

 

Images attached to this report
Displaying report 1-1 of 1.