Reports until 13:19, Sunday 28 September 2025
H1 General
ryan.crouch@LIGO.ORG - posted 13:19, Sunday 28 September 2025 (87186)
OPS Sunday Day shift summary

TITLE: 09/28 Day Shift: 1430-2330 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Observing at 154Mpc
INCOMING OPERATOR: TJ
SHIFT SUMMARY: Quiet shift, we stayed locked the entire time.
LOG: N/A

The LSC front end has an FPU (floating point unit) error. The recent range drop post glitch looks to line up with a ~50 SNR low frequency glitch as seen on the Omicron plot on nuc27, and 2.5 hours ago there's some extra noise in the low frequency SQZ blrms. The recent glitch look to be from  =< 450Hz, as it's seen only in BAND_1-5.

Images attached to this report