SQZ Drops H1 From Observing For About 11min

At the top of the hour (within 15seconds), the Squeezer unlocked and took H1 down.  Tony was in here and thought the timing was odd with it happening close to the top of the hour!  (he saw that the PMC went down at 00:00:15).  

The Log for the SQZ MANAGER also notes the PMC as the cause (w/ a message of "LO railing!").

I slowly got SQZ windows opened & got the wiki up and started recovery by Steps:

Images attached to this report
Comments related to this report
camilla.compton@LIGO.ORG - 08:36, Tuesday 30 September 2025 (87214)

Reason for SQZ unlock was PMC PZT getting to the end of it's range at 0V.

The FC Trans level in the wiki was a red herring (has been corrected), we're unsure why it has dropped so much in the past few months, plot

Images attached to this comment