Took meausrments on SR3 for the L estimator that we're installing.
Settings:
- ISI in ISOLATED
- SUSPOINT matrix set for SR3
- SUS in ALIGNED
- Estimators off (using regular damping)
- L DAMP gain to 20% (-0.1), P DAMP gain at 20% (-0.1), Y DAMP gain to 100% (-0.5)
Results:
SUSPOINT to M1
/ligo/svncommon/SusSVN/sus/trunk/HLTS/H1/SR3/Common/Data/2025-10-07_1800_H1ISIHAM5_ST1_WhiteNoise_SR3SusPoint_EstL_*_0p02to50Hz.xml r12691
M1 to M1
/ligo/svncommon/SusSVN/sus/trunk/HLTS/H1/SR3/SAGM1/Data/2025-10-07_1840_H1SUSSR3_M1_WhiteNoise_EstL_*_0p02to50Hz.xml r12692