richard.mccarthy@LIGO.ORG - posted 14:23, Tuesday 26 November 2013 (8741)
ETM-X SUS cabling shield to ground check.
Tested the cables at the Sat. Amp connector to Ground.  Verifying Pin 13
Was attached to Shield then Pin 13 to Ground.
H1:SUS_ETMX-1 PASSED
H1:SUS_ETMX-4 PASSED
H1:SUS_ETMX-7 PASSED
H1:SUS_ETMX-16 PASSED
H1:SUS_ETMX-19 PASSED
H1:SUS_TMON-1 Failed.  Showed short to ground somewhere between Sat Amp and OSEM.
H1:SUS_TMON-4 PASSED