slow move of OFI temperature
OFI temp [C] OFI PD A power [uW] AS_C_NSUM [W]
37.5 26.5 0.676
33 24.5 0.672
31 24.1 0.672
29 23.8 0.672
25 23.8 0.672
35*  24.9 0.678

I changed the OFI temperature slowly while Robert and Matt were doing other commissoning.  We had only been locked for a bit over an hour when I started, so there is still some thermalization causing a downward drift in AS power, the steps related to OFI temp seem large enough to distinguish. 

For the last point, the thermistor had been at 35C for 30 minutes but temperatures were still moving when we lost lock. 

Images attached to this report
Non-image files attached to this report
Comments related to this report
sheila.dwyer@LIGO.ORG - 14:05, Friday 07 November 2025 (88008)

I've added a script in /ligo/gitcommon/labutils/beam_spot_raster/OFI_temp_stepper.py that would do some temperature steps once an hour.