IMC alignment much better, WFS loops engaged, but IM1 trans light seems clipped in chamber

Stefan and Kiwamu

We continued working on the alignment of IMC today. We managed to get the alignment back i.e. the beam hits the center of the IM4_TRANS QPD with all WFSs in action. It is now happiy locking.

However, we found that the IM1 trans path, which is used for a camera and trigger PD, seemed clipped. By moving the mirrors on IOT2L, we concluded that the clipping happened in chamber, namely in HAM2. This is not a critical issue as this light is used only for triggering purpose and it still gives a reasonable amount of the light. We will keep studying what is happening in the chamber. Perhaps this could be due to the uncontrolled degrees of freedom in the IMC alignment control.

 

What we did:

Comments related to this report
sheila.dwyer@LIGO.ORG - 23:41, Tuesday 03 December 2013 (8808)
This evening Kiwamu and I tried to mitigate the clipping of the im4 Trans beam by putting some angle offsets on mc2 and mc3 while the wfs loops were locked. The hope was that this would change the uncontrolled dofs, while the loops keep the others under control. We used the pico motors to keep the spots centered on the wfs, and had to take small steps. We had to go out to the table and realign the whole path after a few iterations because the power on the lsc pd was dropping. This is a slow process, and didn't fix the problem, the shape of the beam on the imc trans camera did improve. Before and after pics attached
Images attached to this comment
stefan.ballmer@LIGO.ORG - 08:38, Wednesday 04 December 2013 (8809)
Comments on the 5th or 6th phase rotation of the IMC:

We are now using the I demod output for IMC locking. The delay line settings are:
1/16:Up 1/8:Dn 1/4:Up 1/2:Up 1:Up 2:Up 4:Up 8:Dn 16:Dn

The WFS phase settings are:

H1:IMC-WFS_A_SEG1_PHASE_R = -90
H1:IMC-WFS_A_SEG2_PHASE_R = -17
H1:IMC-WFS_A_SEG3_PHASE_R = -27
H1:IMC-WFS_A_SEG4_PHASE_R = -22
H1:IMC-WFS_B_SEG1_PHASE_R = +15
H1:IMC-WFS_B_SEG2_PHASE_R = -34
H1:IMC-WFS_B_SEG3_PHASE_R = +28.5
H1:IMC-WFS_B_SEG4_PHASE_R = -81

For previous rotations, see Oct 1st alog 7941.


kiwamu.izumi@LIGO.ORG - 23:13, Wednesday 04 December 2013 (8827)

There is a similar report from the last May (alog 6265). I wonder if an optic is loose somewhere in this path.