H1 SQZ
sheila.dwyer@LIGO.ORG - posted 08:57, Thursday 11 December 2025 (88474)
SQZ SHG locking

Yesterday we had trouble locking the SHG after the beckhoff restart.  The scan range for the PZT was restored correctly after the restart to 45-100V, but it seems to no longer be the right range to find a resonance.  This morning I changed the range to 0-100V, this allowed us to lock with the PZT voltage around 50.  I've now set it to 20-100V and accepted this in SDF.