Reports until 09:13, Thursday 08 January 2026
H1 IOO (IOO)
masayuki.nakano@LIGO.ORG - posted 09:13, Thursday 08 January 2026 - last comment - 11:21, Tuesday 13 January 2026(88713)
JAC in HAM1

[Jason, Rahul, Jennie, Betsy, Masayuki]

Summary

JM1 was swapped from a tip-tilt suspension to a fixed Siskiyou mount to improve beam stability for alignment, profiling, and JAC control. Mode matching to the JAC was measured using a beam profiler, yielding <1.5% mismatch in both axes, well within the current requirement. The JAC pedestal and body were installed in HAM1, and initial beam injection showed clear HoM resonances at the transmission port. One discrepancy in body mode damper mounting holes was identified and will be tracked via FRS.

Details

 

Images attached to this report
Comments related to this report
jennifer.wright@LIGO.ORG - 09:43, Thursday 08 January 2026 (88715)

I attached the photos of the readout from the nanoscan with the first one being the beam size at the front of the rail, second labelled zero is at zero of the rail, and the others are labelled with the position on the rail in mm.

Images attached to this comment
corey.gray@LIGO.ORG - 11:21, Tuesday 13 January 2026 (88746)EPO

Tagging for EPO photos