david.barker@LIGO.ORG - posted 11:04, Sunday 11 January 2026 - last comment - 08:11, Monday 12 January 2026(88734)
Sun CP1 Fill, 10:00 fill did not run, manually ran a fill at 10:50

Regularly scheduled fill at 10:00 did not run because both thermocouples were above the +30C nominal max limit [31C, 34C] presumably due to icing around the TC and/or Ref Junction.

Sun Jan 11 10:00:00 2026 ALARM: Abort run, Both Thermocouples BAD

I made a temporary change to the configuration to increase the nonimal TC-MAX to +40C and scheduled at 10:50 run.

Sun Jan 11 10:58:34 2026 INFO: Fill completed in 8min 30secs

This appears to be a good fill. The settings will be restored to nominal for tomorrow's fill.

 

Images attached to this report
Comments related to this report
david.barker@LIGO.ORG - 08:11, Monday 12 January 2026 (88736)

Monday's fill will also run with the elevated NOM_MAX=40C settings. Currently TC are [31C,34C].