Displaying report 1-1 of 1.
Reports until 16:16, Wednesday 28 January 2026
H1 CAL
anthony.sanchez@LIGO.ORG - posted 16:16, Wednesday 28 January 2026 (88949)
PCAL TX Module Maintenance.

Yesterday Dripta & I went to the End X Station to do a TX module Maintenance.
The results of this TX Module Maint can be found here T1600436

 

Laser SN01 Incoming laser power from T1300100              
Date 6/4/2013   Aug. 3, 2018 Mar. 19, 2019 Jan-18-2024 April-22-2025 Jan-27-2026  
Laser Shutter Check     PASS PASS Pass Pass Pass  
Max OFS Offset     9.9 V 7.8 V 7.7V 7.7V 7.75  
95% OFS Offset     9.405 V 7.4 V 7.3V 7.315 7.36  
Operating OFS Offset     4.7025 V 3.7 V 3.85V 3.6575 3.68  
Laser Output Power 2.05 W   1.72 W 1.75 W 1.67W 1.7 1.662  
After-Laser Rejected Power     56.5 mW 54 mW 36.6mW 34.6mW 35.7mW  
AOM Input Power     1.60 W 1.69 W 1.59W 1.65 1.64  
Max Diffracted Power     0.967 W 1.13 W 1.1W 1.14 1.1  
Un-Diffracted Power     0.359 W 0.306 W 266mW 0.307W 0.297  
AOM Diffraction Efficiency     60.44% 67% 69.00% 0.690909090909091 0.670731707317073  
After-AOM Rejected Power     12.8 mW 13.7 mW 14.7mW 15.2mW 16.7mW  
TxPD Power     11.1 mW 6.06 mW 6.36mW 11.9mW 6.14mW  
OFSPD Power     5.7 mW 6.56 mW 6.36mW 5.7mW 6.58mW  
Outer Beam Power     228 mW 0.257 W 265mW 0.536 512  
Inner Beam Power     228 mW 0.255 W 260mW 0.526 501  
Output Beam Power Ratio     1 0.992217898832685 0.9811 0.98134328358209 0.978515625  
OFS Gain     38.19 dB 41 dB 39.6db 39.6 39.79  
OFS Phase Margin     60.9 degrees 55 degrees 43deg 42.1 42.3  
ALOG         75484   88949  
                 
Images attached to this report
Non-image files attached to this report
Displaying report 1-1 of 1.