WP13018 Upgrade h1oaf0 18bit-DAC to 20bit-DAC
Fil, Oli, Anamaria, Robert, Ryan S, EJ, Dave:
We replaced the first DAC in h1oaf0 (a 18bit-DAC) with a 20bit-DAC. We reused the IO Chassis ribbon cable and Interface card, they are identical for these types of DAC.
This DAC is only being used by the PEM model.
A new h1iopoaf0 model was installed with the DAC change, its INI file was not changed.
A new h1pemcs model was installed with several changes:
. All 8 DAC channels are now being driven (previously last channel was not driven)
. An excitation stage was added to the model, a copy from LLO's PEM models. This comprises two Oscillators and a Noise_generator (see medm below)
. An additional DACOUTF filter bank as added just prior to the DAC part, each with a x4 filter in the FM10 slot.
A DAQ restart was required.
| 18bit-DAC (removed) | 101208-66 |
| 20bit-DAC (installed, originally in h1susb123) | 210303-49 |
Attached image shows H1PEM_CS_DAC_DRIVES_CUST.adl MEDM which maps the excitation path from OSC/NOISE_GEN through the filters for each DAC channel. Note on DACOUTF, each FM10 has a 20BitDAC filter, which is a x4 gain.
As an example, image shows chan0 being driven by a 0.1Hz sine wave with amp=1.0. The image is caught when the input is 0.988, but the DAC_chan0 drive is 3.953.
Recovery of PSL camera images.
Fil, Corey, Dave:
Following the replacement of camera power supplies, I recovered the PSL cameras and quad video server images.
DAQ Restart
Dave:
The DAQ was restarted for the h1pemcs model change. There were no problems with this restart.
Tue10Feb2026
LOC TIME HOSTNAME MODEL/REBOOT
11:28:08 h1oaf0 ***REBOOT*** <<< Power cycle of h10af0 to upgrade to 20bit-DAC
11:29:51 h1oaf0 h1iopoaf0 <<< new iop model
11:30:04 h1oaf0 h1pemcs <<< new pem model
11:30:17 h1oaf0 h1tcscs
11:30:30 h1oaf0 h1susprocpi
11:30:43 h1oaf0 h1seiproc
11:30:56 h1oaf0 h1oaf
11:31:09 h1oaf0 h1calcs
11:31:22 h1oaf0 h1susproc
11:31:35 h1oaf0 h1calinj
11:31:48 h1oaf0 h1bos
11:38:10 h1daqdc1 [DAQ] <<< 1-leg restart
11:38:22 h1daqfw1 [DAQ]
11:38:23 h1daqtw1 [DAQ]
11:38:24 h1daqnds1 [DAQ]
11:38:31 h1daqgds1 [DAQ]
11:44:32 h1daqgds0 [DAQ] <<< 0-leg restart
11:44:39 h1daqfw0 [DAQ]
11:44:39 h1daqtw0 [DAQ]
11:44:40 h1daqnds0 [DAQ]