Reports until 13:37, Thursday 19 February 2026
LHO VE
jordan.vanosky@LIGO.ORG - posted 13:37, Thursday 19 February 2026 - last comment - 16:07, Friday 06 March 2026(89199)
Leak Checking of the HAM7 +Y door ports and the Relay Tube

Now that HAM7 is at sufficient pressure (6.4E-7 Torr as of this post) we helium leak checked the two re-installed 12" CF blanks removed for table locking and the entire relay tube assembly which had been removed for the viewport adapter on HAM5.

The Helium background was unstable, so we turned on the HAM7 cleanroom (on 2/18 afternoon) to help flush the area of helium overnight.

Today the background stabilized around 1.5E-10 Torr-l/s, so we were able to continue leak checking. No significant helium signal observed above the leak detector background which hovered between 1.5&2E-10 Torr-l/s during leak checking. 

HAM7 continues to pump down with the turbopump, next steps are chamber RGA scans.

Images attached to this report
Comments related to this report
jordan.vanosky@LIGO.ORG - 16:48, Thursday 26 February 2026 (89292)

HAM7 RGA scans and scan information posted to T2600064. RGA is dirty, but the chamber comparison to last pumpdown (Mar 2024) is posted.

gerardo.moreno@LIGO.ORG - 06:32, Friday 27 February 2026 (89296)VE

Late entry.

Last Tuesday, 2/24/2026, HAM7 ion pump was introduced to the chamber's main volume, and currently the chamber is pumped with both; the ion pump and the turbo pump.  Attached is a trend snap of the chamber's pressure since the introduction of the ion pump.

Images attached to this comment
gerardo.moreno@LIGO.ORG - 16:07, Friday 06 March 2026 (89412)VE

Vent Recovery Update

On Wednesday (3/4/26) the turbo pump was isolated from HAM7 to leave the ion pump take care of the pumpdown. 

Since the ion pump was able to take of the pumpdown on its own, on Thursday the SS500 cart then was turned off, and eventually detached from the turbo pump at HAM7.  HAM7 is back to normal operations.

Images attached to this comment