Displaying report 1-1 of 1.
Reports until 13:20, Monday 23 June 2025
H1 TCS (DetChar-Request)
camilla.compton@LIGO.ORG - posted 13:20, Monday 23 June 2025 - last comment - 16:46, Wednesday 25 February 2026(85247)
2Hz Comb in DARM from ITMX HWS Camera?

Ansel, Sheila, Camilla

Last week, Ansel noticed that there is a 2Hz comb in DARM since the break, similar to that that we've seen from the HWS camera sync frequency and power supplies and fixed in 75876.  The cabling has not been changed since, the camera sync frequency has been changed.

Our current camera sync frequencies are: ITMX = 2Hz, ITMY = 10Hz. We have typically seen these combs in H1:PEM-CS_MAG_LVEA_OUTPUTOPTICS_Y_DQ. With a 0.0005Hz BW on DTT I can't easily see these combs, see attached.

Images attached to this report
Comments related to this report
evan.goetz@LIGO.ORG - 13:31, Monday 23 June 2025 (85254)DetChar
It may be difficult to see in a standard spectrum, but can be clearly seen in Fscan plots linked off of the summary pages. For the "observing" Fscan, the interactive spectrum plot shows the 2 Hz comb marked automatically. See the attached image of H1:GDS-CALIB_STRAIN_CLEAN
Images attached to this comment
camilla.compton@LIGO.ORG - 17:04, Tuesday 24 June 2025 (85306)

Verifed that the cabling has not changed since 75876.

Next steps we should follow, as listed in 75876 would be to try using a different power supply or lowering the voltage to +12V. Or, there is a note suggesting Fil could make a new cable to power both the camera and CLink's via the external supply (14V is fine for both). 

evan.goetz@LIGO.ORG - 18:30, Tuesday 24 June 2025 (85312)DetChar
Thanks Camilla. If anything can be done more rapidly than waiting another week, it would be very much appreciated. Continuing to collect contaminated data is bad for CW searches.
camilla.compton@LIGO.ORG - 15:11, Wednesday 25 June 2025 (85341)

Matt and I turned down the Voltage supplied from 14V to 12V for each camera at ~22:00UTC when the IFO was relocking. Verified HWS cameras and code still running.

We also will plan to have Dave reimpliemnt the hws_camera_control.py script he wrote in 74951 to turn the HWS's off in Observing until we fix this issue.

kiet.pham@LIGO.ORG - 11:58, Thursday 26 June 2025 (85361)

The 2 Hz comb is still present in H1:GDS-CALIB_STRAIN_CLEAN after the voltage change (before the software update)

Images attached to this comment
camilla.compton@LIGO.ORG - 16:46, Wednesday 25 February 2026 (89272)

alog 85678 showed this 2Hz comb appeared to not be the HWS camera. 

Displaying report 1-1 of 1.