HAM1 SUS TFs now that we're in vacuum

Took transfer functions for the suspensions in HAM1 since they're now in a state where they're not being affected by air anymore. They're all looking good. RM1 and RM2 are flipped in phase, but they've been like that (84543).

Settings:
- ISI DAMPED
- DAMP OFF
- OPTICALIGN OFFSETS OFF

JM1
Data:
/ligo/svncommon/SusSVN/sus/trunk/HTTS/H1/JM1/SAGM1/Data/2026-03-03_1730_H1SUSJM1_M1_WhiteNoise_{L,P,Y}_0p02to50Hz.xml
Results:
/ligo/svncommon/SusSVN/sus/trunk/HTTS/H1/JM1/Results/2026-03-03_1730_H1SUSJM1_M1_ALL_TFs.pdf
r12929

JM3
Data:
/ligo/svncommon/SusSVN/sus/trunk/HTTS/H1/JM3/SAGM1/Data/2026-03-03_1800_H1SUSJM3_M1_WhiteNoise_{L,P,Y}_0p02to50Hz.xml
Results:
/ligo/svncommon/SusSVN/sus/trunk/HTTS/H1/JM3/Results/2026-03-03_1800_H1SUSJM3_M1_ALL_TFs.pdf
r12930

PM1
Data:
/ligo/svncommon/SusSVN/sus/trunk/HTTS/H1/PM1/SAGM1/Data/2026-03-03_1840_H1SUSPM1_M1_WhiteNoise_{L,P,Y}_0p02to50Hz.xml
Results:

/ligo/svncommon/SusSVN/sus/trunk/HTTS/H1/PM1/Results/2026-03-03_1840_H1SUSPM1_M1_ALL_TFs.pdf
r12931

RM1
Data:
/ligo/svncommon/SusSVN/sus/trunk/HTTS/H1/RM1/SAGM1/Data/2026-03-03_1855_H1SUSRM1_M1_WhiteNoise_{L,P,Y}_0p02to50Hz.xml
Results:
/ligo/svncommon/SusSVN/sus/trunk/HTTS/H1/RM1/Results/2026-03-03_1855_H1SUSRM1_M1_ALL_TFs.pdf
r12933

RM2
Data:
/ligo/svncommon/SusSVN/sus/trunk/HTTS/H1/RM2/SAGM1/Data/2026-03-03_1910_H1SUSRM2_M1_WhiteNoise_{L,P,Y}_0p02to50Hz.xml
Results:
/ligo/svncommon/SusSVN/sus/trunk/HTTS/H1/RM2/Results/2026-03-03_1910_H1SUSRM2_M1_ALL_TFs.pdf
r12935

Comparison
Results:
/ligo/svncommon/SusSVN/sus/trunk/HTTS/Common/Data/allhttss_2026-03-03_HAM1_BackinVac_ALL_TFs.pdf
r12936