Per WP13076 we replaced the coil driver for MC1 due to suspicion it was causing MC1 to jump. The coil driver in question glitched on TOP3. Referencing D0902810-v12, the chassis in question is U41 on SUS-C4, we exchanged it with a spare.
Old SN - S1001097, New SN - S1100194
Shelia will retest the signal chain to see if this solves the issue.
It did not, at least with comparing a before/after transfer function. See alog89409.