[Sheila, Jennie, Jenne, RyanS, Corey]
We have been able to get PRMI locked (with arms held off resonance using ALS). To do this we:
PRMI locked without too much fuss. Since we've effectively also increased the noise on the error signals, it's a bit delicate, and I lost lock twice while touching up alignment. The third lock I was able to go a bit slower when touching up BS and PRM alignment, and we've now been locked for more than 20 mins (aka infinity). Jennie is taking PRCL and MICH OLGs.
The attached image shows all the things that I had changed.
Nothing is in guardian (except maybe some trigger thresholds which should be reverted), and nothing is saved in SDF.
I used the lscparams to change the MICH triggers so I shave put these values back.
Below are OLG TFs I ran from userapps/lsc/h1/templates/PRCL and userapps/lsc/h1/templates/MICH
For PRCL the measurement is shown here where blue is the reference measurement in the tenplate, and red and green are the measurements I took today with different injection amplitudes.
We had to raise the injection in PRCL by a factor of 2.5 ti check the wiggles we can see in the gain between 10 to 30Hz are real.
Here is the measurement for MICH. Not sure we understand either of these yet. More investigation to come tomorrow.
Here is an ndscope with the 'good PRMI ACQUIRE time' we were using to scale gains and trigger levels. That is a prmi where we eventually locked and I picked the highest flash that occurred before we did.