Displaying report 1-1 of 1.
Reports until 17:04, Friday 13 March 2026
H1 ISC
jennifer.wright@LIGO.ORG - posted 17:04, Friday 13 March 2026 (89503)
POPAIR and REFLAIR checks

Jennie W, Sheila D,

 

I compared the RF monitor values for POPAIR and REFLAIR LSC photodioes between today during state 51 "PRMI LOCKED" and 2026-11-14 22:28:00 UTC in state 50 "ACQUIRE PRMI".

Note we re-aligned both POPAIR B and REFLAIR A yesterday.

POPAIR A RF9 and RF45 have not changed much.

POPAIR B RF18 and RF90 have reduced a lot.

As have REFLAIR A RF9 and RF45.

REFLAIR B RF27 and RF135 have gone up instead of down which seems less worrying for locking purposes.

Name 14th Nov[dBm] Today[dBm] P_today / P_Nov
POPAIR_A_RF9 -54.8

-54.9

0.98
POPAIR_A_RF45 -54.6 -55 0.91
POPAIR_B_RF18 -49.6 -53.9 0.37
POPAIR_B_RF90 -46.9 -51.4 0.35
REFLAIR_A_RF9 -45.4

-55.4

0.1
REFLAIR_A_RF45 -37.5 -54.6 0.02
REFLAIR_B_RF27 -53.1 -50.0 2.0
REFLAIR_B_RF135 -30.2 -28.0 1.7

 

After this Oli and Sheila re-aligned REFLAIR A (detailed in Sheila's log 89501).