H1 SUS (SUS)
rahul.kumar@LIGO.ORG - posted 15:26, Monday 30 March 2026 - last comment - 15:28, Monday 30 March 2026(89702)
OSEM spectra for SUS PR2 and PR3 - looks healthy

Sheila, Rahul

This morning I took top mass osem spectra for both PR2 and PR3 suspension and they looks healthy (the noise level for all bosems are similar and the magnitude is to below 10^-4  above 10Hz - un-calibrated) - plot attached below. There is a 15Hz peak seen on PR3, which is also seen (in both PR2 and PR3) if we go back in time (7 days) compare the spectra results. 

I also looked at the medm screen of both the suspensions (input filters, flag position, damping loops and coil driver settings) and trended bosem channels and did not find anything unusual. 

Images attached to this report
Comments related to this report
rahul.kumar@LIGO.ORG - 15:28, Monday 30 March 2026 (89704)

spectra for PR2 and PR3 from last week (25 March 2026). 

Images attached to this comment