Displaying report 1-1 of 1.
Reports until 16:19, Wednesday 01 April 2026
H1 SPI (SPI)
sina.koehlenbeck@LIGO.ORG - posted 16:19, Wednesday 01 April 2026 (89739)
LIGO-SPI Photodetector Chassis: CDS Range Maximization and TIA Output Settings

Sina, Jeff

The photodetector chassis of the LIGO-SPI pathfinder has been modified to maximize the CDS range. The TIA is designed to put out 0V - 10V on the positive leg of the differential output. The power monitoring PDs are set to get close to 5V, the interferometer PDs a mean of 2.5V, and the QPD quadrants 3.5V each. The whitening filters for the heterodyne-interferometer photodiodes are bypassed. The details of the installed capacitor and resistors are listed in the following table:

Detector R4 Resistor value now (kohm) Design value of pole frequency (Hz) C5 capacitor value calculated (F) C5 capacitor value now (pF) Resulting pole frequency (Hz) Chassis Variant Channel Mean power at PD (mW) measured with S121C Mean voltage at PD positive leg (V) Sum of mean voltage of TIA positive leg (mV) PD calibration (mW/V)
PD POW: FBR_PWRIN_MEAS (SIG) 53 1.00E+04 3.00E-10 220 13,650 3 1 0.9 3.9 3.9 0.23
PD POW: FBR_PWRIN_REF (LO) 53 1.00E+04 3.00E-10 220 13,650 3 2 1 4.25 4.25 0.24
                       
QPDB               9.9 13.5 13.5 0.73
QPDB Q1 3 1.00E+04 5.31E-09 1,000 53,052 2 5   2.9 2.9  
QPDB Q2 3 1.00E+04 5.31E-09 1,000 53,052 2 6   3 3  
QPDB Q3 3 1.00E+04 5.31E-09 1,000 53,052 2 7   3.9 3.9  
QPDB Q4 3 1.00E+04 5.31E-09 1,000 53,052 2 8   3.7 3.7  
                       
PD IFO: IFO_MEAS_A (LO)               3.3 0.9    
PD IFO: IFO_MEAS_A (SIG) 3 5.00E+04 1.06E-09 1,000 53,052 2 3 4.8 1.29 2.19 3.7
PD IFO: IFO_MEAS_B (LO)               3.1 0.9    
PD IFO: IFO_MEAS_B (SIG) 3 5.00E+04 1.06E-09 1,000 53,052 2 4 4.8 1.36 2.26 3.5
                       
PD IFO: IFO_REF_A (LO)               2.6 1.16    
PD IFO: IFO_REF_A (SIG) 4.99 5.00E+04 6.38E-10 560 56,955 2 1 2.9 1.24 2.4 2.29
PD IFO: IFO_REF_B (LO)               2.6 1.08    
PD IFO: IFO_REF_B (SIG) 4.99 5.00E+04 6.38E-10 560 56,955 2 2 2.9 1.22 2.3 2.39
                       
QPDA 1.5 10000 1.06E-08         18.5 9 9 2.06
QPDA Q1 1.5 10000 1.06E-08 2000 53,052 1 1   1.9 1.9  
QPDA Q2 1.5 10000 1.06E-08 2000 53,052 1 2   2.6 2.6  
QPDA Q3 1.5 10000 1.06E-08 2000 53,052 1 3   2.2 2.2  
QPDA Q4 1.5 10000 1.06E-08 2000 53,052 1 4   2.3 2.3  

 

The capacitor for the QPDs might be changed to result in a low-pass frequency of 10 kHz at a later date.

Displaying report 1-1 of 1.