PCAL EY Endstation measurement

Dripta and I went to EY with PS4 to do a PCAL ES measurement. We followed the  T1500062-v21 Procedure  but there were changes to the scripts we were using to do the analysis for this measurement. 
Obligitory before beam spot

python3 generate_measurement_data.py --WS PS4 --date 2026-04-02
Reading in config file from python file in scripts
../../../Common/O4PSparams.yaml
PS4 rho, kappa, u_rel on 2026-04-02 corrected to ES temperature 299.9 K :
-4.7013350503199405 -0.0002694340454223 0.0007506545774987483
Copying the scripts into tD directory...
Connected to h1daqnds1
martel run
reading data at start_time:  1459616830
reading data at start_time:  1459617300
reading data at start_time:  1459617635
reading data at start_time:  1459618001
reading data at start_time:  1459618370
reading data at start_time:  1459618700
reading data at start_time:  1459619030
reading data at start_time:  1459619580
 reading data at start_time:  1459619930
Ratios: -0.5341166097396405 -0.5439653195648569
writing nds2 data to files
finishing writing
Background Values:
bg1 =        18.656956; Background of TX when WS is at TX
bg2 =        5.029697; Background of WS when WS is at TX
bg3 =        18.678297; Background of TX when WS is at RX
bg4 =        5.152238; Background of WS when WS is at RX
bg5 =        18.675415; Background of TX
bg6 =        -0.527308; Background of RX

The uncertainty reported below are Relative Standard Deviation in percent 

Intermediate Ratios
RatioWS_TX_it      = -0.534117;
RatioWS_TX_ot      = -0.543965;
RatioWS_TX_ir      = -0.526729;
RatioWS_TX_or      = -0.535441;
RatioWS_TX_it_unc  = 0.057495;
RatioWS_TX_ot_unc  = 0.055038;
RatioWS_TX_ir_unc  = 0.052784;
RatioWS_TX_or_unc  = 0.054928;
Optical Efficiency
OE_Inner_beam                      = 0.986428;
OE_Outer_beam                      = 0.984304;
Weighted_Optical_Efficiency        = 0.985366;

OE_Inner_beam_unc                  = 0.042212;
OE_Outer_beam_unc                  = 0.043766; 
Weighted_Optical_Efficiency_unc    = 0.060806;

Martel Voltage fit:
Gradient      = 1637.917034;
Intercept     = 0.063908;


 Power Imbalance = 0.981895;

Endstation Power sensors to WS ratios::
Ratio_WS_TX                        = -0.927573;
Ratio_WS_RX                        = -1.384310;

Ratio_WS_TX_unc                    = 0.046237;
Ratio_WS_RX_unc                    = 0.042014;

=============================================================
============= Values for Force Coefficients =================
=============================================================

Key Pcal Values :
GS           =      -5.135100; Gold Standard Value in (V/W)             
WS           =      -4.701335; Working Standard Value             

costheta     =      0.988362; Angle of incidence
c            =      299792458.000000; Speed of Light
             
End Station Values : 
TXWS         =        -0.927573; Tx to WS Rel responsivity (V/V)
sigma_TXWS   =        0.000429; Uncertainity of Tx to WS Rel responsivity (V/V)
RXWS         =        -1.384310; Rx to WS Rel responsivity (V/V)
sigma_RXWS   =        0.000582; Uncertainity of Rx to WS Rel responsivity (V/V)

e            =        0.985366; Optical Efficiency
sigma_e      =        0.000599; Uncertainity in Optical Efficiency

Martel Voltage fit : 
Martel_gradient         =        1637.917034; Martel to output channel (C/V)
Martel_intercept   =        0.063908; Intercept of fit of     Martel to output (C/V)

Power Loss Apportion : 
beta          =        0.998844; Ratio between input and output (Beta)  
E_T          =        0.992082; TX Optical efficiency 
sigma_E_T          =        0.000302; Uncertainity in TX Optical efficiency 
E_R          =        0.993230; RX Optical Efficiency 
sigma_E_R          =        0.000302; Uncertainity in RX Optical efficiency 

Force Coefficients : 
FC_TxPD          =        9.158224e-13; TxPD Force Coefficient 
FC_RxPD          =        6.227715e-13; RxPD Force Coefficient 
sigma_FC_TxPD          =        4.284257e-24; TxPD Force Coefficient 
sigma_FC_RxPD          =        2.800786e-24; RxPD Force Coefficient 
data written ... to /ligo/gitcommon/Calibration/pcal/O4/ES/measurements/LHO_EndY/tD20260407
 


Martel_Voltage_test.png 
WS_at_RX.png 
WS_at_TX.png 
WS_at_RX_BOTH_BEAMS.png 
Final Report is here: LHO_EndY_PD_ReportV6.pdf
Final beamspot .

 

Images attached to this report
Non-image files attached to this report