Reports until 09:02, Thursday 23 April 2026
H1 SUS (OpsInfo)
ryan.short@LIGO.ORG - posted 09:02, Thursday 23 April 2026 (89983)
BS OpLev Damping

Since we've found the BS OpLev damping on and causing the suspension to be moving around more than we'd like during prep for the swap to the BBSS, I went through Guardian nodes looking for all the places the OpLev damping is either turned on or off as a learning exercise for myself:

On/Off Node State Method
ON SUS_BS ALIGNED All inputs and outputs on
ON ISC_DRMI DOWN Inputs and outputs on, gains to 300,650 (P,Y)
OFF ISC_DRMI ENGAGE_DRMI_ASC Gains ramp to zero (only if using MICH ASC)
OFF ALIGN_IFO MICH_DARK_ALIGN

Gains ramp to zero

ON ALIGN_IFO MICH_DARK_OFFLOADED Gains to 300,650 (P,Y)

Basically, the BS OpLev damping is ON during alignment and lock acquisition, and is turned OFF once MICH ASC is engaged. To keep the OpLev damping off for the time being, I've set their gains to zero since the suspension node doesn't change those and we won't be going through any states that do for a while.

When the BBSS is installed, we should think about how if at all these interactions need to change (although I'm sure someone already has).