Displaying report 1-1 of 1.
Reports until 15:48, Thursday 23 April 2026
H1 IOO (ISC)
jennifer.wright@LIGO.ORG - posted 15:48, Thursday 23 April 2026 - last comment - 17:36, Tuesday 28 April 2026(89988)
Comparison with IMC WFS before and after JAC installation

Jennie W, Jeff Kissel,

 

Summary: Can we see a difference in jitter with JAC in the IMC ASC sensors? Short answer, yes.

The IMC has less angular motion with JAC installed at some frequencies - jitter is being suppressed by the JAC so less input beam motion relative to the cavity axis. The QPD readouts are noisier with JAC but there is one new suspension between JAC and the IMC so maybe the beam pointing to MC1 is noisier? MC2 TRANS QPD looks pretty similar with and without JAC, but this makes sense as cavity axis is still well aligned to this QPD. I need some other figure of merit to look at jitter as it only limits the IFO above 20 Hz, see alog #86555 by Sheila.

Since there was no time for a full IFO lock betwen the HAM1 vent and the current vent of the corner volume, we cannot directly check whether the jitter coupling to DARM is better with the JAC cavity.

We have lots of data of the JAC and IMC locked after the installation of JAC, however.

So I compared two times when we had roughly 2W input from the PSL to HAM1. In both cases H1 and the corner were at vacuum to rule out confounding effects from purge air and other sound noise seen when HAM1 was at air.

In both these cases the IMC was locked.

The two times I used were:

2025/11/17 15:38:54 UTC no JAC, during O4c

2026/03/20 00:26:01 UTC JAC installed, HAM1 at vacuum

Here is the ASD measurements for five sensors in the pitch degree of freedom:

IMC REFL WFS A  Roughly the same above 10Hz with and without JAC. Better with JAC between 0.3 and 7Hz.

IMC REFL WFS B Roughly the same above 10Hz with and without JAC. Better with JAC below 7Hz.

IMC REFL QPD A has lower noise now below 5Hz than before JAC installation and much higher noise at frequencies above 10Hz.

IMC REFL QPD B has lower noise now below 3Hz than before JAC installation and much higher noise at frequencies above 10Hz.

MC2 TRANS QPD roughly the same with and without JAC.

Jitter noise is more of a problem in the 20-800 Hz region, which these plots don't tell us much about.

I also need the check YAW

 

 

 

Images attached to this report
Comments related to this report
jennifer.wright@LIGO.ORG - 16:54, Thursday 23 April 2026 (89992)

I also plotted the yaw DOFs for each of the sensors I used above and for the same two reference times.

Images attached to this comment
jennifer.wright@LIGO.ORG - 17:36, Tuesday 28 April 2026 (90048)

I think it is impossible to tell from the measurements above whether the jitter is better. The peaks that could be jitter in the QPD measurements (WFS_A_DC and WFS_B_DC) are hard to compare as the background noise level is larger for the March measurements (with JAC). Even though the input power was roughly the same I suspect that the total SUM value of the QPDs was different between these two measurement times. Since pitch and yaw are normalised by the total sum value that would explain why the two WFS QPD measurements have different background noise levels. I can't see any obvious jitter peaks in the demodulated WFS signals (WFS_A_I and WFS_B_I).

Displaying report 1-1 of 1.