To check for any rubbing or other issues that the BSFM might be encountering on the test stand, I took some transfer functions today after Betsy plugged the top mass in yesterday (90063). The measurements have really poor coherence because the BSFM isn't covered, but it's decent enough to at least verify that TFs are looking good and we aren't near touching anywhere.
Data
$(sustrunk)/BSFM/H1/BS/SAGM1/Data/2026-04-30_1615_H1SUSBS_M1_WhiteNoise_{L,T,V,R,P,Y}_0p02to50Hz.xml