SQZ beam profiles with different psams

Ryan Short, Tony Sanchez, Sheila

We took beam profiles of the sqz beam in the homodyne path using the Thorlabs M2Ms beamprofiler extension.  It made collecting data for a variety of psams settings much faster than using a scanning slit profiler alone.  

Tips for using this very nice extender kit:

We spent some time refinding the IR beam on the SQZT7 IR PD, documented this in it's own alog so it's easy to search: 90183

We set the profiler as shown in the attached photo.   The flipper mirror is 50.75" from the bottom persicope mirror, a steering mirror is 225mm from that, and the reference plane is 260mm from the steering mirror, so the reference plane is 1.774 meters from the bottom periscope mirror.  Leo Schrader has a helpful list of distances in the google document linked to T2500228.

 

 

Images attached to this report
Non-image files attached to this report
Comments related to this report
sheila.dwyer@LIGO.ORG - 16:34, Monday 11 May 2026 (90201)

Keita, Ryan Short, Tony, Sheila

In the saved PDFs, the units on "Beam Waist Position" are listed as um, but in the text file (which does contain all the same information), the values are the same but the units are mm.  mm must be the correct units. 

Looking at T2500077  we believe that to get from the reported Beam Waist position in the text file we need to add (z_stage_min - 4.4 mm= 122.9mm) to the distance from the reference plane as depicted in that diagram to the waist position.  We believe that this waist position being positive means that the waist is after the reference plane (on the sensor side of the reference plane), I think a negative number would mean the waist is before the profiler.  

I asked Claude to write a script to collect the data from the txt files in the archive above and save it in a yml, the script is attached here in case someone else wants it. 

 

Non-image files attached to this comment
sheila.dwyer@LIGO.ORG - 14:00, Tuesday 26 May 2026 (90344)

I had further correspondence with Thorlabs tech support about understanding what the "beam waist position" is relative to.  I believe that for the "original waist position" a positive number means the waist is closer to the source, and is reported in mm from the reference plane. the after lens measurement seems to have the opposite sign, and I am not sure how the value given relates to the reference plane. 

 