H1 SUS (SUS)
ryan.crouch@LIGO.ORG - posted 10:42, Thursday 14 May 2026 (90245)
OPLEV charge measurements, ETMX, ETMY

I ran the oplev charge measurements this morning for both of the ETMs, previously checked in alog89906.

ETMY's charge remains largely unchanged, the values across the quadrants remain below 100 on all quadrants/DOF and I don't see any appreciable trends. As a reminder we switched off the HV digitally for EY.

ETMX's charge is decreasing across all quadrants/DOF after the sign flip we inposed back in April. Over the past 4 weeks the average decrease for P was ~ -14 [V], and ~ -18.25 [V] for Y. UL saw the greatest decrease, a drop of over 30 [V] for both P & Y, UR saw the smallest decrease, around ~ 5 [V] for both P & Y. The charge is now at or below 50 [V] on all quadrants/DOF except UR_P and LR_P.