Reports until 16:21, Monday 18 May 2026
H1 SPI
jeffrey.kissel@LIGO.ORG - posted 16:21, Monday 18 May 2026 - last comment - 10:50, Tuesday 19 May 2026(90270)
2026-05-18 SPI Pathfinder Status Update
J. Kissel

Picking up from last Friday (2026-05-15, LHO:90253), today I:
    - Secured D_R_P1 and D_M_P1 and confirmed that TFP REFL beams went into the dumps
    - Started aligning REF path and migrating optics over from the D2400107-v4 breadbord's CVM100 mounts into / onto D2400107-v5 IXM100 mounts:
        . R_M1 migrated.
        . Re-set up the power meter gantry assembly (since its only me, and I need the power meter stable for maximizing throughput thru irises)
        . Placed irises in holes 83 and 84, adjusted R_F1 and R_M1 to walk and align the beam thru irises, confirming with IR camera view of iris holes and power meter thru-put. 
        . Secured R_F1 and R_M1 IXM mount 8-100 alignment screw shaft set screws, since these optics shall not move again.
        . R_B1 migrated.
        . Aligned beam through holes 86 and 95, repeating process above, but only adjusting R_B1.
        . Uncapped IFO_REF A and B PDs, PWRIN_REF PD and hooked them up to the o-scope (discovering a minor re-cabling issue; see LHO:90268)
            UNSOLVED SIDEQUEST (1) :: PWRIN_REF PD has a ~120 Hz oscillation on it.
        . R_B2 migrated.
        . Aligned beam through holes 98 and 99, repeating process above.
            UNSOLVED SIDEQUEST (2) :: Now, with D_FBR_PWRIN_REF so far away, the known issue with its PD seating within the generic enclosure means the PD reflected beam is ~2 inches above the board. Tried a huge 2"x3" dump but that was too big.
        . R_B3 migrated.
        . With nothing to align, just checked that beam went through irises in holes 100 and 97, and that beam looked well centered on PD. They did.
        . Migrated D_IFO_REF_B and D_IFO_REF_A, confirmed PD reflection landed on there and it did.

Done for the day! The MEAS PATH alignment tomorrow!
Images attached to this report
Comments related to this report
jeffrey.kissel@LIGO.ORG - 10:35, Tuesday 19 May 2026 (90277)
Regarding UNSOLVED SIDEQUEST (2) :: Now, with D_FBR_PWRIN_REF further away [...] the PD reflected beam is ~2 inches above the board.

Here're some picture to aide the discussion.

I say "this is a known issue," because we'd encountered this particular PD's pitch of the diode when assembling the CVM100 version of the ISIK transceiver, D2400107-v4, but in that layout, the PD was angled away from the incoming beam in the opposite direction and the the dump was closer. We didn't have room behind R_B2 to have the same dump location with the upgrade to IXM100 mounts.

None of the other PDs have this issue, the beam reflects back at the same 1 [inch] beam height that's the beam height for the board.

We were in too much of a rush and had too little experience/expertise with these to consider cracking open the Generic PD enclosure (D1600083) to try to fix it.
We suspect, in this Type 3 assembly for the FFD200 PDs, that either 
    - the PD (line item 12 T1000573) is not seated well in its retaining ring (line item 4 D1600082),   
    - there's a defect or something stuck between the retaining and its seat in the enclosure housing (line item 1 D1600079)
    - the retaining ring is not *exactly* the same size as the Q3000 for which it was designed and there's some slop in the assembly, or
    - the PD's pins were soldered into the PCB at an angle (line item 11 D1700116)
But we're just guessing. 

This is S2401094, whose assembly technique is discussed in CIT:837 and record of completion is mentioned in CIT:898.

Starting up the conversation with CIT.
Images attached to this comment
jeffrey.kissel@LIGO.ORG - 10:50, Tuesday 19 May 2026 (90279)
Regarding UNSOLVED SIDEQUEST (1) :: PWRIN_REF PD has a ~120 Hz oscillation on it.

See attached o-scope picture -- channel 3 (purple). 2V peak-to-peak wobbly 120 Hz oscillation on a 4.5 [V] mean.

We've not tested this PD since we updated its transimpedance circuitry (see LHO:90105, and TIA Variant 3 D1002481 instantiation S2500713), so there may be a grounding issue there. Hard to believe it's the TIA op-amp itself, given that the RC feedback impedance we installed yields a 13.497 [kHz] pole.

I've only tested the IFO MEAS and IFO REF channels of the Variant 2 chassis with light so far, but they don't show any issue.

Laser power level on this PD during yesterday's test is the expected ~0.8 [mW].

I checked (but didn't re- or triple-check) that the electrical ground situation is well-managed. As in March 2026, the electrical ground of power supply feeding the TIAs banana'd-to-BNC'd-to-clip-doodled to the shield the o-scope's clip-doodle dongle whose signal is reading out PD ADC output voltage from a breakout board. Nowhere along this adpater/doodle/breakout/cable chain is metal touching metal.

I'll look through the chain again, try reading it out differentially, and consult with local experts. It isn't a high priority right now, but we need to understand this before the transceiver leaves the optics lab.
Images attached to this comment