Displaying report 1-1 of 1.
Reports until 08:43, Thursday 04 June 2026
H1 SEI
arnaud.pele@LIGO.ORG - posted 08:43, Thursday 04 June 2026 - last comment - 10:24, Thursday 04 June 2026(90476)
CRS Report June 3 - 1st CRS suspended

Shoshana, Alexandra, Jim, Arnaud, Fil

Yesterday's activity log: 

  Nominal New Fork (D2300094-v4) New Fork + D230099 type2 (4g vs 2g type 1) 
Z +186um +189.8um +229.2um
X -2.4um -3.75um -3.75um
Images attached to this report
Comments related to this report
alexandra.mitchell@LIGO.ORG - 10:24, Thursday 04 June 2026 (90482)
The monitor channels in the CRS pre-amp D2500390 are connected up incorrectly. 
The following table details the current connections:

HoQI/TEST POINT CONNECTION - MON CHANNEL CONNECTION 
HA PDS - HD PDS
HA PDC - HA PDC
HA PDMC - HD PDMC

HB PDS - HC PDS
HB PDC - HB PDC
HB PDMC - HC PDMC

HC PDS - HB PDS
HC PDC - HC PDC
HC PDMC - HB PDMC

HD PDS - HA PDS
HD PDC - HD PDC
HD PDMC - HA PDMC
Displaying report 1-1 of 1.