Noticed the drive to ISI actuators gettting larger and JeffK thought it might be that the Boost2 was on with the level3 controllers. It was nearing ACtuator trigger levels and we switched it to boost3. Maybe too late or maybe you must start from scratch but it still tripped the Stage2 WDs. Jeff brought it back up to level3 and still in boost2 rather than 3. We had though it might have been related to me just jumping staight to level3 from level1 earlier. Attempting to run a spectra to see what is happening but my session had died onme twice--I'll take this up later.
Meanwhile--maybe there is a script error engaging level2 boost in the level3 isolation? For the moment I'll leave it in level1.