Displaying report 1-1 of 1.
Reports until 13:01, Tuesday 09 June 2026
H1 COC
anamaria.effler@LIGO.ORG - posted 13:01, Tuesday 09 June 2026 - last comment - 08:07, Wednesday 10 June 2026(90550)
ITMY first contact pour complete

Ibrahim, Ryan, Betsy, Anamaria

This morning we poured the first contact onto ITMY. All went well. We had 3 minor drops at the bottom, but they dried fast and didn't continue leaking. We finished pouring about 09:35, took some 10 min. We then kept pushing the pico motors until the level stopped changing and the picos started to bottom out. 

It took us a bit to get the camera setup with reasonable lighting, so for next time: we should turn off the illuminator from the getgo and set up a tripod just for light, next to the camera. (Because of the CP, we are looking through 3 surfaces to see the HR of the test mass so it's quite hard to not have all the reflections cover what we really want to monitor.)

Comments related to this report
betsy.weaver@LIGO.ORG - 08:07, Wednesday 10 June 2026 (90560)
Images attached to this comment
Displaying report 1-1 of 1.