Reports until 16:32, Wednesday 17 June 2026
H1 SUS (SUS)
ibrahim.abouelfettouh@LIGO.ORG - posted 16:32, Wednesday 17 June 2026 - last comment - 17:21, Wednesday 17 June 2026(90656)
BBSS Gross Alignment Efforts

Ibrahim, Betsy, Keita

The BBSS Oplevs aren't to be trusted (alog 90648). 

Today, we went into BSC2 to attempt to use ITMX and ITMY as a lever using the PSL beam. We put in 500mW flashes with the rotation stage locked out.

We found the beam nicely (multiple modes) on both ITMs, albeit a bit yawed. We moved PR3 from -233 to -583 in Yaw, which recovered the beam on the center of both ITMX and ITMY.

For context, we had already centered the BBSS to the structure on the test stand but found that we were not on the oplev once we went back into the cartridge.

We're now working on seeing if we can trust this BBSS alignment, or if we have to make mechanical or offset changes.

Comments related to this report
keita.kawabe@LIGO.ORG - 17:21, Wednesday 17 June 2026 (90661)

At the end of the day, PSL waveplate rotator was set to 200mW and then denergized and locked, and the light pipe was closed.

I made the following changes for the whitening and didn't change them back. I will, later.

  Used to be Temporarily changed for now
ASC-AS_C 18dB, 2 whitening stages 18dB, 3 whitening stages
ASC-OMC_A and B 0dB, 2 whitening stages 27dB, 3 whitening stages
ASC-POP_A and B 12dB, no whitening 9dB, 2 whitening stages

ASC-AS_A_DC and B gain were high and I didn't change them.