Shoshana, Jim, Arnaud
The x6 HAM3 new CPS boards have been set to the following configuration (as per E2200455):
| board SN | probe SN | Location | |
| H1 | 29536 | 22777 | Slot 1 in crate +Y side |
| V1 | 38690 | 22781 | Slot 2 in crate +Y side |
| H3 | 12563 | 22779 | Slot 3 in crate +Y side |
| V3 | 38494 | 22775 | Slot 4 in crate +Y side |
| H2 | 23370 | 22778 | Slot 1 in crate -Y side |
| V2 | 11950 | 22780 | Slot 2 in crate -Y side |
There was some glitching noticed on the channels that were in range prior to this work, which was fixed after this setting change (see attached). Two of the cards were set to Master before and Jim is confident this would have created problems.
Left to do :
1- Label the cards at the front of the boards with module SN and corner #
2- Disconnect the grounding to the chamber of the crate on the +Y side
3- Label the in-air lemo cables with H1, V1, H2 ...
4- Update ICS with the new boards SN
5- Gather the cards that were removed from WHAM3 to send back to microsense for recalibration with a new set of sensors (RMA)
The tracking spreadsheet for this ECR was updated here: E2500184