[Jim, Shoshana]
The CRS is currently aligned, suspended, balanced, and unlocked on HAM3!
I aligned each HoQI as well as I could on the table, HoQI2 is slightly better than HoQI1, but it shouldn't be an issue. I also lowered the resonant frequency as much as possible, but I ended up running out of mass that I could easily remove from the bottom, so right now the resonant frequency is at ~25Hz. It's possible this raised frequency is because these flexures (21, 22) might be slightly thicker than the previous ones as they came from a different batch.
Tested the picomotor, moving the X value positive moves the mass to the left (towards HoQI1). We had it do three revolutions in both directions and the picomotor was able to couple with the mass adjuster while at maximum range (tilting fully to either side). We've also discovered that the GS13s outputs go crazy when the picomotor is running, which is something to be aware of.
We've verified that voltage makes it to the damping capacitor plates on either side. We've also confirmed that in the HOQI{1,2}_DIST data increases as the corner cube gets closer to either interferometer (i.e. when the left corner cube (wing) gets closer to HoQI1, the HoQI1_DIST channel data increases)
Attached are various photos of the CRS on the table for future reference, as well as a guide to all the balance masses currently on it.
Notes on CRS commissioning tonight :
The overnight measurement is slightly better, but we are missing coherence below 15mHz which is probably due to too much air currents. The measurement is still good enough to fit and allow us to extract some physical parameters. Attached are updated data, scripts and the latest fit.
The zpk coefficients for this fit are :
zpk([0.0009 + 0.0766i 0.0009 - 0.0766i],[-0.0005 + 0.1584i -0.0005 - 0.1584i],1.0262)
My fitting script agrees. Looks like a delta value of 10 um.