This morning, with Mitch and Shoshana, I went into the nozzle between HAM3 and BSC2, via the HAM3 door to fix the broken H3 actuator. I had to remove a couple baffles to get in and get access to the hatch over the H3 actuator, once the hatch was off, I found the pin on the back of the actuator terminal block had come lose, probably when I was in there to do the CPS upgrade. I probably should have checked this wire after I finished that swap, but it there was no way to see that anything was wrong, I had to grab the wire and pull on it. Once that was reconnected and I put everything back together, we did a quick chamberside test and H3 was driving again. I've now completed close out tfs, attached image, and run a CPS linearity measurement to check the new cpses. The linearity test plots failed in a way that didn't save the data, but I do have the terminal output of the test which agrees with HAM7:
'Actuator H1 to CPS H1'
Linear regression y= 1.862 x + 245
'Actuator V1 to CPS V1'
Linear regression y= 1.890 x + -771
'Actuator H2 to CPS H2'
Linear regression y= 1.845 x + -271
'Actuator V2 to CPS V2'
Linear regression y= 5.105 x + 2771
'Actuator H3 to CPS H3'
Linear regression y= 5.073 x + 1543
'Actuator V3 to CPS V3'
Linear regression y= 5.119 x + 2882
At some point after doors go on, I will want to try to fix the cps linearity test script and run it again, among the many other measurements I will need for HAM3.
Arnaud helped me get Huyen's update of this script, which worked after restarting a stale matlab session. Linearity of the new cpses still look good, slopes of cps vs actuator output are all consistent in a way that makes sense.
'Actuator H1 to CPS H1'
Linear regression y= 1.865 x + 225
'Actuator V1 to CPS V1'
Linear regression y= 1.883 x + -803
'Actuator H2 to CPS H2'
Linear regression y= 1.849 x + -316
'Actuator V2 to CPS V2'
Linear regression y= 5.521 x + 3170
'Actuator H3 to CPS H3'
Linear regression y= 5.470 x + 1383
'Actuator V3 to CPS V3'
Linear regression y= 5.556 x + 3076
I will try to get long tfs of the ISI this weekend, to see if loops need touching up.