Shoshana, Arnaud
wp 13430
CRS laser was turned back on. HOQI signals are alive, and CRS is still ringing down at 25mHz (fig 1)
Some convention we implementend :
1) hoqi photodiode signals calibrated in mW with calibration gain (see below) and filter was added to FM1 of the HOQI{1,2}_{SIN,COS,MCOS} filterbanks. A minus sign was added to have positive power at the output of the hoqi filterbanks (fig 2)
[1] PD sensitivity
[2] Transimpendance Amplfier Gain
[3] 16 bit ADC
2) hoqi distance was calibrated as per section 2 in the assembly procedure such that a positive hoqi distance signal corresponds to the retroreflector on the CRS moving away from the sensor. For this, we had to add a -1 gain in the HOQI{1,2}_DIST filterbank (fig 3)
The next two conventions not yet implemented are to match the sign of the CRS with the CPS and get the CRS tilt signal calibrated in nrad (at the CRSINF filterbank)