Reports until 09:57, Monday 20 July 2026
LHO VE (VE)
travis.sadecki@LIGO.ORG - posted 09:57, Monday 20 July 2026 - last comment - 14:25, Friday 24 July 2026(91121)
HAM5/7 relay tube viewport adapter installed

This is a late entry from last Friday.  We have removed the relay tube between HAM5 and HAM7 and replaced it with the adapter/viewport assembly on the HAM5 side.  The assembly is being pumped down independently by a small turbo/aux cart.  The GV should not be opened until the pressure in the viewport assembly is approximately the same as the corner pressure and the aux cart is valved out.

Comments related to this report
gerardo.moreno@LIGO.ORG - 14:25, Friday 24 July 2026 (91239)VE

The adapter's isolation valve was opened, then the adapter was leak checked, no leaks were found.  Adapter and accessories will remain until we are given the OK to remove them.

Images attached to this comment