The first image is sdf diffs created after we phased the JAC WFS readouts this morning, the next two are cdsFilt banks that have no filters and a gain 1 but need to be on to route signals from the wavefront sensor readout matrixes I_mtrx and Q_mtrx to the normalised channels H1:JAC-WFS_A_I_PIT_NORM, H1:JAC-WFS_A_I_YAW_NORM, H1:JAC-WFS_B_I_PIT_NORM, H1:JAC-WFS_B_I_YAW_NORM. Confusingly, these are not displayed on the standard 'JAC_WFS_Settings' screens so I might have to put them on there so this doesn't happen in future (this is probably some legacy infrastructure I copied over from the IMC_WFS model blocks that we don't really need).