BBSS In-vac OLGs taken

I took a set of OLGs yesterday for the BBSS now that it's in vac. I was just comparing them to the last BSFM ones that we have from 2023 (in blue), which obviously will look a bit different.

Settings
- BS in HEALTH_CHECK but with:
    - DAMP ON (you need damping on to run OLGs)
- HEPI and ISI are in their nominal ISOLATED states (ISI no ST2 boost)

Data
/ligo/svncommon/SusSVN/sus/trunk/BBSS/H1/BS/SAGM1/Data/2026-07-20_2200_OLG_tfs/2026-07-20_2300_H1SUSBS_M1_CDBIOState_1_WhiteNoise_{L,T,V,R,P,Y}_0p01to50Hz_OpenLoopGainTF.xml
r13075

Images attached to this report