H1 General
anthony.sanchez@LIGO.ORG - posted 17:45, Friday 24 July 2026 (91247)
Friday Ops shift report.

TITLE: 07/24 Day Shift: 1430-2330 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Planned Engineering
INCOMING OPERATOR: None
SHIFT SUMMARY:
Comissioners got in today and we started trying to lock PRMI and align the input arm.

After a short while.... something happened. and there were some suggestions that a power outage had happened. 
Which also coincided with FW0 connection issue. ( The control room was running on FW1 and otherwise the CDS "one leg" as opposed to the zero leg which is good, cause FW0 is not connected.)

So we all started trying to gather data of all the things that would have been tripped or otherwise upset by a power glitch.
We even made a list!: 
    [Sheila] 10:20 PDT JAC REFL power drops -- i.e. we lost the beam out the JAC into the IFO.
    [Sheila] 10:20 PDT JM1 OSEMs show sharp change in behaviour " exponential droop" at the ~few [urad] level (JM1 is upstream of JAC)
    [Sheila] 10:20 PDT IMC PWR IN PD also shows a (faster) "exponential droop" from 2 [W] to 0.8 [W]
    [Sheila] 10:27 PDT Started making changes to laser power guardian (requesting 10 [W]) to get back to 2 [W]. This turns the IO rotation stage. OSEMs on JMs, RM1, seem to witness this
    [Dave] 10:20 PDT h1sush12 DAC2, channel 20 (the JM1 UL channel) shows a spike glitch, at the 0.01e6 [ct] ~= 1.5 [mV] output voltage level (28-bit DAC, so [DAC V] = [DAC ct] * 20/2^28 [DAC V/DAC ct])
    [Dave] 10:20 PDT ADC3 channels 0-3 (JM1's OSEM PDs) change mean behavior, similar "exponential" but up,
    [Dave] 10:29 PDT another sharp "glitch" feature change in the ADC3 CH 0-3 (JM1 OSEMs)
    [Dave] 10:29 PDT Reported "glitch" in the corner station power
    [Dave] 10:33 Frame writer went unstable

So I started looking for the MAINS channels to document the voltage change.

H0:FMC-EX_MAINS_CHAN_1_VOLTAGE looked very strange, so did 2, and 3. They haven't reported Voltage correctly since JUN 18th! Perhaps from a Hardware Swap or a Beckhoff change? 

Thankfully dave has these channels in an ndscope template that still work to determine power glitches:
H1:PEM-CS_MAINSMON_EBAY_1_DQ  , and ebay 2 and 3.

Turns out there was a button pusher and an alog about how  that button no longer exists which you can read about here: alog 91246 . 

M6.0 Earthquake from Vanuatu Shook up End X so much I had to go down to EX and tickle the  H1:HPI-PUMP_EX_RMT_PRESS_PUMP_CTRL_OUTPUT  control loop until it worked again. 

LOG:                                                                                                                                                  

Start Time System Name Location Lazer_Haz Task Time End
14:47 FAC Kim, Dawn H2, MY, EY, MX, EX, HAM SHAQ N Tour of VEA areas and resupplies. 16:13
15:41 VAC Jordan MY N Setting up outside pump 16:59
15:51 PEM Robert Shrey LVEA y Setting up shaker and accelerometers 16:18
16:27 FAC Kim LVEA y Technical Cleaning & resupply 16:55
16:31 PEM Robert , Carlos, Shrey LVEA y Debugging Shaker issues 17:25
17:00 VAC Mitchell & Jordan LVEA y Taking pump cart Down from platform 17:14
17:26 PEM Robert, Carols, Shrey Just over the over pass N Seismometer testing 19:37
18:14 CDS Dave Remote N Restarting the Entire \ 18:44
18:17 ISC Jennie LVEA yes Opening IOT1 to check beams 18:40
18:59 EE Jennie W & Marc CER N Checking on the SAT AMP JM1 19:44
19:37 CDS Jonathan Remote N Restarting 0 leg 19:44
20:24 JAC Sheila, Jennie, Keita LVEA yes Working on the JAC Table 21:09
20:27 PEM Carlos & Miranda Overpass N Setting up seismometers 22:08
20:31 vac Jordan LVEA N Corner RGA work 22:08
22:06 Tour Robert, Shrey & Co LVEA & Roof y Giving a tour of the LVEA 23:48
23:01 ISC Sheila & Jennie LVEA yes Working in ISCT1 23:17
23:35 VAC Jordan Mid Y N Shutting down Pump 23:54
00:40 SEI Tony EX N restarting the HEPI Pump controller. 00:42


 

Images attached to this report