Reports until 10:42, Monday 03 August 2026
H1 CDS
david.barker@LIGO.ORG - posted 10:42, Monday 03 August 2026 (91370)
Greened up h1iopomc0 timing card status on CDS Overview

Post upgrade the CDS Overview was showing an issue with h1iopomc0's timing card. On further investigation we found that this timing card is a very early rollout of the V5 firmware, upgraded back in Feb 2025 for DT-frequency changes. The actual issue is that RCG5.65 is more accurately calculating the duotone crossing for h1omc0's low-noise ADCs. This value was +0.1uS pre-upgrade and now is -5.625uS with only nano-second variation. The CDS overview was using the DT crossing time as an indicator of timing card firmware version. V3 had DTs between +7uS and +8uS, V5's are in the -1uS to +1uS range.

I've changed the CDS Overview to extend the lower limit of acceptible DT timings to -6uS. No further action is required.