Displaying report 1-1 of 1.
Reports until 15:44, Monday 03 August 2026
H1 SQZ
camilla.compton@LIGO.ORG - posted 15:44, Monday 03 August 2026 (91371)
SQZ OMC Scans and M^2 Measurements with new ZM4 ZM5 PSAMS Preloadings

Sheila, Ryan S, Camilla

We spent the morning repeating 90783, now that ZM4 preloading has been adjusted (90951) and ZM5 has been swapped and now has a different preloading than our O4 ZM5 (91292):

Sheila then used her script to run OMC scans at a grid (below) of PSAMS locations. These initially look great, showing sqz to omc mode mismatch is as good as 1%. with some alignment fluctuations. Ndscope of data taking attached. 

PSAM_VALUES_V5 = [-8, -5.5, -3.3, -1]
PSAM_VALUES_V4 = [-3.2, -1, 1.2, 3.4] 

Ryan then took M^2 measurements on SQZT7 of a slightly different grid:

PSAM_VALUES_V5 = [-8,  -6.75, -5.5, -4.5, -3.3, -1]
PSAM_VALUES_V4 = [-3.2, -1, 1.2, 3.4] 
Images attached to this report
Displaying report 1-1 of 1.