PCAL EY End Station Measurement

Ibrahim and I went to End Y with the PS4 Working Standard to do a PCAL End Station measurement.  We followed T1500062-V21 to complete the measurment suite with out any issues.
Here are the obligitory Before Beam spot pics.  
Martel plots for ADC conversion.

Working standard at TX module with only One beam. 

Working standard at the RX module with only One beam.

Working standard at RX module with Both beams.

Post measurement beam spots!



anthony.sanchez@cdsws25: python generate_measurement_data.py --WS PS4 --date 2026-07-16
Reading in config file from python file in scripts
../../../Common/O4PSparams.yaml
PS4 rho, kappa, u_rel on 2026-07-16 corrected to ES temperature 299.6 K :
-4.698647654693516 -0.0002694340454223 0.000750622815531554
Copying the scripts into tD directory...
Connected to h1daqnds1
martel run
reading data at start_time:  1469911700
reading data at start_time:  1469912270
reading data at start_time:  1469912670
reading data at start_time:  1469913100
reading data at start_time:  1469913500
reading data at start_time:  1469913900
reading data at start_time:  1469914560
reading data at start_time:  1469915320
reading data at start_time:  1469915650
Ratios: -0.5339859682199335 -0.543765733929218
writing nds2 data to files
finishing writing
Background Values:
bg1 =        18.671164; Background of TX when WS is at TX
bg2 =        5.156755; Background of WS when WS is at TX
bg3 =        18.605483; Background of TX when WS is at RX
bg4 =        5.281252; Background of WS when WS is at RX
bg5 =        18.674174; Background of TX
bg6 =        -1.117857; Background of RX

The uncertainty reported below are Relative Standard Deviation in percent 

Intermediate Ratios
RatioWS_TX_it      = -0.533986;
RatioWS_TX_ot      = -0.543766;
RatioWS_TX_ir      = -0.526609;
RatioWS_TX_or      = -0.535095;
RatioWS_TX_it_unc  = 0.059001;
RatioWS_TX_ot_unc  = 0.057064;
RatioWS_TX_ir_unc  = 0.061781;
RatioWS_TX_or_unc  = 0.063032;
Optical Efficiency
OE_Inner_beam                      = 0.985936;
OE_Outer_beam                      = 0.983757;
Weighted_Optical_Efficiency        = 0.984846;

OE_Inner_beam_unc                  = 0.046465;
OE_Outer_beam_unc                  = 0.047412;
Weighted_Optical_Efficiency_unc    = 0.066385;

Martel Voltage fit:
Gradient      = 1637.889101;
Intercept     = 0.085727;


 Power Imbalance = 0.982015;

Endstation Power sensors to WS ratios::
Ratio_WS_TX                        = -0.927858;
Ratio_WS_RX                        = -1.383422;

Ratio_WS_TX_unc                    = 0.047658;
Ratio_WS_RX_unc                    = 0.041564;

=============================================================
============= Values for Force Coefficients =================
=============================================================

Key Pcal Values :
GS           =      -5.135100; Gold Standard Value in (V/W)             
WS           =      -4.698648; Working Standard Value             

costheta     =      0.988362; Angle of incidence
c            =      299792458.000000; Speed of Light
             
End Station Values : 
TXWS         =        -0.927858; Tx to WS Rel responsivity (V/V)
sigma_TXWS   =        0.000442; Uncertainity of Tx to WS Rel responsivity (V/V)
RXWS         =        -1.383422; Rx to WS Rel responsivity (V/V)
sigma_RXWS   =        0.000575; Uncertainity of Rx to WS Rel responsivity (V/V)

e            =        0.984846; Optical Efficiency
sigma_e      =        0.000654; Uncertainity in Optical Efficiency

Martel Voltage fit : 
Martel_gradient         =        1637.889101; Martel to output channel (C/V)
Martel_intercept   =        0.085727; Intercept of fit of     Martel to output (C/V)

Power Loss Apportion : 
beta          =        0.998844; Ratio between input and output (Beta)  
E_T          =        0.991820; TX Optical efficiency 
sigma_E_T          =        0.000329; Uncertainity in TX Optical efficiency 
E_R          =        0.992968; RX Optical Efficiency 
sigma_E_R          =        0.000330; Uncertainity in RX Optical efficiency 

Force Coefficients : 
FC_TxPD          =        9.158396e-13; TxPD Force Coefficient 
FC_RxPD          =        6.237025e-13; RxPD Force Coefficient 
sigma_FC_TxPD          =        4.417816e-24; TxPD Force Coefficient 
sigma_FC_RxPD          =        2.847865e-24; RxPD Force Coefficient 
data written to ../../measurements/LHO_EndY/tD20260804/
 

 