Displaying report 1-1 of 1.
Reports until 14:17, Friday 07 August 2026
H1 CDS
marc.pirello@LIGO.ORG - posted 14:17, Friday 07 August 2026 (91448)
CER Kepco Power Supplies

WP13495

The following supplies were replaced in the CER Mezanine.

Kepco Power Supply in VDD-C6 Rack, Left Slot U31-29, ISC-R3&R5 +24V
Old S12001945
New S12001992

Kepco Power Supply in VDD-C6 Rack, Right Slot U31-29, ISC-R3&R5 -24V
Old S1201946
New S1201947

The negative supply was oscillating between 22V and 21V when measured before replacing, possibly due to failed fan.  New supplies were adjusted to +/-24V at the racks.

Caroline C, Marc P.

Displaying report 1-1 of 1.