Displaying report 1-1 of 1.
Reports until 16:12, Friday 07 August 2026
H1 SQZ
sheila.dwyer@LIGO.ORG - posted 16:12, Friday 07 August 2026 - last comment - 15:30, Saturday 08 August 2026(91444)
HAM7 power budget repeat

Ryan S, Sheila

In summary, after removal of the A:L2 apperture stop we still see about 4% extra losses in HAM7.  We searched for the rejected beam from SFI1 in both directions with lights out, IR card, and viewer.  To explain a few % loss we should expect to have 10-20urad in this beam, which should be visible with lights out an a viewer.  We also noticed that the beam is off in yaw at the location where we had an iris after ZM1.  

Since Ryan and Camilla aligned the beam onto the SQZT7 PD yesterday, we are noting the power on that PD here after each measurement so we can use that to account for fluctuations in power out of the OPO.  

From the control room, we dithered ZM2 + ZM3 and tried to reduce the fluctuations in the SQZT7 IR PD, we were able to recover a 4% of power this way, with fairly small adjustments (+9urad ZM2 P, +19 urad ZM2 yaw, 5 urad ZM3 yaw).  

One more round of power measurements: 

 

Images attached to this report
Comments related to this report
ryan.short@LIGO.ORG - 17:19, Friday 07 August 2026 (91454)SQZ

After all of this power budgeting, OPOS rebalancing, ZM4 PSAM preloading adjustments, and alignment, this afternoon I took another (possibly the last?) ZM4/5 PSAM M^2 grid on SQZT7. Data files attached. I used different strain gauge values for ZM4 than in the past since they've shifted with the preloading change, but this still covers pretty much the whole PZT range.

This was a 26-point grid using the following PSAM strain gauge settings:

ZM4 PSAMS = [-2.6, -1.0, 1.0, 4.0]

ZM5 PSAMS = [-8.0, -6.75, -5.5, -4.5, -3.3, -1.0]

Non-image files attached to this comment
sheila.dwyer@LIGO.ORG - 15:30, Saturday 08 August 2026 (91455)

The attached plot shows the qs measured on SQZT7 propagated to SRM, with three different ZM4 preloads. Camilla has summarized the history of our ZM4 + ZM5 preloading in 91404

 

Clickable Image

The blue points here are with 75 in lbs on ZM4, as adjusted in 75677.  The teal points were after an attempt to lower the pre-load to 65 in lbs, which Rahul and Camilla later rechecked with the torque wrench and thought that it went to something between 40-60 in lbs. ( 91416)

On Wensday, Rahul and Camilla believed that they applied more torque to the psams (91416), but somehow the beam profiles that Ryan took show that the preload was reduced instead.  

We did not repeat the OMC scans at this preload.  The M^2 range is similar to previous measurements, but the astigmatism is a little better here.  

Images attached to this comment
Displaying report 1-1 of 1.