8/11/26
[Huyen, Jim, Michael, Shoshana]
CRS Commissioning (from list in G2601523):
Noise Hunting:
Continued Noise Hunting
Results:
Next Steps:
Put the HoQI signals through a high pass in the front end, but this didn't fix the noise issue. (picture 1)
We grounded the damping pins to see if that would reduce some of the noise.
Additionally, the TEC should be set to 10.72, and stay there as that is ideal for the laser
Damped down the CRS as much as it could go and took a spectra and left the damping on. When the damping is left on and soon after it is turned off, it looks like there is some slight improvement at the 1-3Hz range in the SUM measurement (picture 1).
Additionally, we want to look at the drift that CRS sees, it's probably the CRS still settling, but it's something to keep an eye on