Displaying report 1-1 of 1.
Reports until 15:57, Tuesday 11 August 2026
H1 SQZ (SUS)
ryan.short@LIGO.ORG - posted 15:57, Tuesday 11 August 2026 - last comment - 10:26, Thursday 13 August 2026(91484)
ZM4 PSAMs Preload Adjusted (Again, Final)

R. Kumar, S. Dwyer, C. Compton, R. Short

Following the work done yesterday adjusting the preload on the ZM4 PSAMs (alog91471), Rahul and I set about adjusting it further today to get our calculated projected beam size on SRM close to where we want it. Using the same procedure as yesterday, I turned off the ZM4/5 PSAM drive chassis, Rahul went into HAM7, locked down ZM4, torqued the preload about 1/8" of a turn (it's hard to quantify a torque spec in this manner), then finally unlocked the suspension. I moved the ZM4 pitch alignment slider about -1400 counts to realign the beam on SQZT7 to our alignment irises as a result of this torque adjustment. I then proceeded to take a couple of beam profiles on the table with the M^2 device to see where we landed. This plot [initial] shows the difference between before and after this adjustment in terms of the projected beam at SRM with a few different PSAM settings on ZM4 and ZM5. The four points I took here informed us that we could stand to torque ZM4 a bit further to hopefully line up with the center of the plot.

Rahul went back into HAM7 and torqued another 1/8" and I took another couple of profiles after (I did not need to move ZM4's alignment sliders after this adjustment). I took a total of nine profile points here, all of which can be seen compared to other measurements today on this plot [full]. Since there are points in the bullseye, we decided to stop here and that this is where the ZM4 preloading will stay. This plot [final] shows just the beam profiles taken at our final preloading position.

Tomorrow we will move on to final closeout checks in HAM7.

Images attached to this report
Comments related to this report
ryan.crouch@LIGO.ORG - 16:26, Tuesday 11 August 2026 (91486)SUS

I ran a health check TF on ZM4 and it looks good.

2026-08-11_2200_H1SUSZM4_M1_WhiteNoise_L_0p02to50Hz.xml
2026-08-11_2200_H1SUSZM4_M1_WhiteNoise_P_0p02to50Hz.xml
2026-08-11_2200_H1SUSZM4_M1_WhiteNoise_Y_0p02to50Hz.xml

Non-image files attached to this comment
rahul.kumar@LIGO.ORG - 10:26, Thursday 13 August 2026 (91515)

The current pre-load value on the ZM4 is 60in-lb + 1/8th + 1/8th turn using a 0.5in size wrench.

Displaying report 1-1 of 1.