Displaying report 1-1 of 1.
Reports until 16:50, Sunday 16 August 2026
H1 IOO (IOO)
masayuki.nakano@LIGO.ORG - posted 16:50, Sunday 16 August 2026 (91558)
JAC pico motor calibration by non-locked sweeps.

All four JAC pico axes (heads A and B, X and Y motors) have been calibrated in counts to beam displacement on the WFS, by non-locked sweeps read through the WFS DC signals. Efficiencies come out at 0.6–1.2e-3 beam radii per count (0.09–0.19 μm, 0.8–1.6 μrad of beam pointing), with a direction dependence of 9–27 %.


Setup


Conversion

The fit gives the efficiency in beam radii (d/w) per count. Beam radii at the QPDs from the JAC reflected-path optical model:

plane wx [μm] wy [μm]
WFS A 199.4 181.0
WFS B 150.0 155.9

Results

Denominator: SYS-MOTION_C_PICO_D_CURRENT_X/Y_POSITION [counts] of the swept motor. "up" / "down" refer to the sign of the count sweep.

pico / motor QPD axis direction d/w per count μm per count μrad per count (beam)
A / X A YAW (x) up -9.41e-4 0.188 1.64
A / X A YAW (x) down -7.82e-4 0.156 1.36
A / Y A PIT (y) up -7.33e-4 0.133 1.16
A / Y A PIT (y) down -8.05e-4 0.146 1.27
B / X B YAW (x) up -1.22e-3 0.183 1.60
B / X B YAW (x) down -9.32e-4 0.140 1.22
B / Y B PIT (y) up -5.72e-4 0.089 0.78
B / Y B PIT (y) down -7.00e-4 0.109 0.95

Attachments: per-axis sweep figures (normalized signal against pico counts).

Images attached to this report
Displaying report 1-1 of 1.